Planejamento fatorial e superfície de resposta: otimização de um método voltamétrico para a determinação de ag(i) empregando um eletrodo de pasta de nanotubos de carbono
AUTOR(ES)
Vicentini, Fernando Campanhã, Figueiredo-Filho, Luiz Carlos Soares, Janegitz, Bruno C., Santiago, Aline, Pereira-Filho, Edenir Rodrigues, Fatibello-Filho, Orlando
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2011
RESUMO
A factorial design applied in a voltammetric stripping method for the measurement of Ag(I) in natural water is described. The procedure is based on the effective pre-concentration of silver ions on electrode surface. The calibration graph was linear in the silver concentration range from 7.92 x 10"7 to 1.07 x 10"5 mol L"1 with a detection limit of 3.81 x 10-7 mol L-1. The determination of Ag(I) in natural water samples was carried out satisfactory with the proposed electrode.