Comparação da resistencia mecanica a flexão da fixação interna rigida atraves do uso de parafusos metalicos e reabsorviveis na osteotomia sagital do ramo mandibular : estudo in vitro
AUTOR(ES)
Rubens Guimarães Filho
DATA DE PUBLICAÇÃO
2003
RESUMO
The purpose of this study was to evaluate the flexion biomechanical resistance of two internal rigid fixation systems. The study which were mechanically tested in twenty sheep s hemimandibles, divided in two groups of ten in each. All mandibles were subjected to a ramus sagittal split osteotomy, followed by an advancement of 5 mm. The samples were measured to avoid discrepancies between them. Group A were fixed by three reabsorbable screws of 2.0 X 12 mm (Bionx Implants®, Tampere, Finland), inserted in positional way and disposed in inverted "L". The same happened in-group B, which were used metallic screws (Osteomed® CO. Addison, Texas - USA), with the same dimensions and design. The results showed no significant statistical difference between the groups. The reabsorbable system tested proved to be a viable alternative when compared to the metallic one
ASSUNTO(S)
odontologia osteotomia cirurgia
ACESSO AO ARTIGO
http://libdigi.unicamp.br/document/?code=vtls000293697Documentos Relacionados
- Avaliação da resistência à flexão da fixação interna rígida, com parafusos absorvíveis e metálicos, na osteotomia sagital do ramo mandibular: estudo in vitro
- Avaliação da resistencia a flexão da fixação interna, utilizando parafusos absorviveis e metalicos, na osteotomia sagital do ramo em hemimandibulas de carneiros : estudo in vitro
- Estudo cefalometrico da estabilidade do recuo mandibular atraves da osteotomia sagital dos ramos mandibulares e fixação interna rigida
- Estudo quimico, macroscopico e da resistencia a flexão de placas e parafusos de titanio usados na fixação interna rigida
- Estudo químico, macroscópico e da resistência à flexão de placas e parafusos de titânio usados na fixação interna rígida