Validation of a Radioimmunoassay for Indole-3-Acetic Acid Using Gas Chromatography-Selected Ion Monitoring-Mass Spectrometry 1
AUTOR(ES)
Pengelly, William L.
RESUMO
A radioimmunoassay for indole-3-acetic acid (IAA) has been validated by comparison with a physico-chemical assay utilizing gas chromatography-selected ion monitoring-mass spectrometry and 4,5,6,7-tetradeutero-indole-3-acetic acid as an internal standard. The radioimmunoassay provided a reliable estimate of the free IAA content of etiolated corn shoots. However, base hydrolysis of extracts for determination of ester IAA released substances which interfered with the radioimmunoassay. Interference was detected by internal controls and by lack of agreement with the mass spectral method. Interfering compounds could be removed from extracts by chromatography on diethylaminoethyl- and hydroxypropylated (lipophilic) Sephadex G-25. Following such purification the radioimmunoassay estimate of the total (free + ester) IAA content of etiolated corn shoots agreed with the mass spectral method within 2% on the average.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=425896Documentos Relacionados
- Comparison of a Commercial ELISA Assay for Indole-3-Acetic Acid at Several Stages of Purification and Analysis by Gas Chromatography-Selected Ion Monitoring-Mass Spectrometry Using a 13C6-Labeled Internal Standard 1
- A Selective Gas Chromatography-Selected Ion Monitoring Mass Spectrometry (GC-MS/SIM) Method to Determine Polychlorinated Biphenyls in Fuller’s Earth
- Simultaneous Quantitation of Indole 3-Acetic Acid and Abscisic Acid in Small Samples of Plant Tissue by Gas Chromatography/Mass Spectrometry/Selected Ion Monitoring
- Indole-3-acetic Acid in Douglas Fir: Analysis by Gas-Liquid Chromatography and Mass Spectrometry
- A Microscale Technique for Gas Chromatography-Mass Spectrometry Measurements of Picogram Amounts of Indole-3-Acetic Acid in Plant Tissues.