Use of BACTEC 9240 Blood Culture System for Detection of Brucella melitensis in Synovial Fluid
AUTOR(ES)
Yagupsky, Pablo
FONTE
American Society for Microbiology
RESUMO
Synovial fluid specimens aspirated from patients with arthritis were inoculated into an aerobic Peds Plus blood culture bottle and monitored by the BACTEC 9240 instrument for 4 weeks. A total of 1,072 synovial fluid cultures were processed, and 15 (0.14%) were positive for Brucella melitensis. A single culture, harboring 1.3 CFU of viable organisms per ml, was missed by the instrument. All 14 positive BACTEC cultures were detected within 3 to 7 days.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=87808Documentos Relacionados
- Detection of Brucella melitensis by BACTEC NR660 blood culture system.
- Use of the BACTEC 9240 System with Mycosis-IC/F Blood Culture Bottles for Detection of Fungemia
- Use of the Isolator 1.5 Microbial Tube for Detection of Brucella melitensis in Synovial Fluid
- Comparison of BACTEC 9240 Peds Plus medium and isolator 1.5 microbial tube for detection of Brucella melitensis from blood cultures.