Modelo de calibração da concentração de metilmetacrilato em solução aquosa utilizando espectroscopia de absorção no ultravioleta
AUTOR(ES)
Cunha Jr., Albano Porto da, Cunha, Vicente de Paula Prisco da, Silveira Jr., Landulfo, Martin, Aírton Abrahão
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2003-12
RESUMO
Dilutions of methylmetacrylate ranging between 1 and 50 ppm were obtained from a stock solution of 1 ml of monomer in 100 ml of deionised water, and were analyzed by an absorption spectrophotometer in the UV-visible. Absorbance values were used to develop a calibration model based on the PLS, with the aim to determine new sample concentrations. The number of latent variables used was 6, with the standard errors of calibration and prediction found to be 0,048 ml/100 ml and 0,058 ml/100 ml. The calibration model was successfully used to calculate the concentration of monomer released in water, where complete dentures were kept for one hour after polymerization.