Evaluation of rubella immune status by three commercial enzyme-linked immunosorbent assays.
AUTOR(ES)
Field, P R
RESUMO
Three commercial indirect enzyme-linked immunosorbent assays (ELISAs) (Enzygnost-Rubella, RUBELISA, and ORTHO Rubella) were evaluated for the determination of immune status by testing 1,090 serum specimens, 410 of which were from nonimmune patients. In comparison with the standard reference technique, the hemagglutination inhibition (HAI) test, the sensitivities of ORTHO Rubella (100%) and Enzygnost-Rubella (99.26%) were excellent, whereas the sensitivity of RUBELISA (95.60%) was marginally lower because of the inability of this assay to detect antibody in 22% of the serum specimens with HAI titers of 10 and 11% of sera with HAI titers of 20. The specificity of all three systems was greater than 97%. There was a linear correlation between mean ELISA values and increasing HAI titers (r greater than or equal to 0.94). Both ORTHO Rubella and Enzygnost-Rubella were shown to be suitable replacements for the HAI test, provided that an equivocal zone is incorporated in the ORTHO system and only unheated sera are used in the Enzygnost system.