Ethylene Oxide Gaseous Sterilization: I. Concentration and Temperature Effects

AUTOR(ES)
RESUMO

The relationships of reaction temperature and concentration of gaseous ethylene oxide to the time required for inactivation of air-dried Bacillus subtilis var. niger spores are more complex than previously reported. A plot of temperature vs. the logarithm of “thermochemical death time” (TCDT) resulted in a straight line between 18 and 57 C for systems of “high” ethylene oxide concentration. The TCDT values were independent of ethylene oxide concentrations above certain temperature-dependent limits. A given ethylene oxide concentration produced a TCDT curve identical in the upper temperature regions with that for higher concentrations. As the temperature was lowered beyond a critical point, this curve diverged from that for higher concentrations, as a straight line of lesser slope. Thus, a series of curves exists for a range of ethylene oxide concentrations. They are characterized by two segments, both logarithmic, intersecting at a critical temperature for each concentration. The intersecting point is at a temperature inversely related to the ethylene oxide gas concentration. The temperature quotient for the high temperature segments of all systems was 1.8. This value was characteristic for ethylene oxide concentrations of 440 and 880 mg/liter at temperatures above 40.6 and 33.4 C, respectively. Below these critical temperatures, the Q10 values for the respective systems were 3.2 and 2.3.