Estudo da resistencia a corrosão da liga Ti-GAI-7Nb para aplicação em implantes ortopedicos
AUTOR(ES)
Sirlei da Rosa
DATA DE PUBLICAÇÃO
1997
RESUMO
This work presents a study on the resistance to corrosion ofthe orthopaedic Ti-6AI-7Nb alloy previously, hot annealled and isothermically hot forged, in Hank s solution. Potentiodinamic technique was used for determination of corrosion potential, passivation potential, current of corrosion in the active region and current of passivation. Data recorded on these experiments were used for to establisch a direct relationship between corrosion resistence of the alloy and the microestructure obtained for different conditions of hot forging such as temperature, deformation depth and rate of deformation.
ASSUNTO(S)
biocompatibilidade implantes ortopedicos ligas de titanio eletroquimica forjamento microestrutura materiais resistentes a corrosão
ACESSO AO ARTIGO
http://libdigi.unicamp.br/document/?code=vtls000118368Documentos Relacionados
- Obtenção e caracterização da liga Ti-6AI-7Nb para aplicação em proteses ortopedicas
- Obtenção e caracterização de ligas Ti-Al-Nb para uso em implantes ortopedicos
- Avaliação da resistência à corrosão em aços inoxidáveis ASTM F138 empregados em implantes ortopédicos
- Estudo das propriedades mecânicas da liga Ti-Nb sinterizada para aplicação na área biomédica
- Liga Ti-35Nb-5Ta-7Zr : microestrutura e propriedades mecanicas e de corrosão