Determinação de metil-etil-cetona em amostras de urina com amostragem por micro extração em fase sólida (MEFS) em headspace associada à cromatografia gasosa com detector de ionização de chama (CG-DIC)

AUTOR(ES)
FONTE

Química Nova

DATA DE PUBLICAÇÃO

2008

RESUMO

Methyl ethyl ketone (MEK) is a solvent commonly used in chemical, paint and shoe industry. The aim of this study was to develop and validate a method for urinary quantification of MEK, employing headspace solid phase micro extraction sampling (SPME) coupled to gas chromatography with flame ionization detection (GC-FID). The calibration curve (y=4.6851x-0.0011) presented good linearity with r²=0.9993. Accuracy (94-109%), intra-assay precision (4.07-5.91%) and inter-assay precision (3.03-5.62%) were acceptable. The quantification limit was 0.19 mg/L. This low cost method can be used routinely in the biological monitoring of occupational exposure to MEK, according to the requirements of the Brazilian legislation.

Documentos Relacionados