Determinação de manganês em material particulado atmosférico de ambientes de trabalho utilizando eletrodo de diamante dopado com boro e voltametria de onda quadrada com redissolução catódica

AUTOR(ES)
FONTE

Química Nova

DATA DE PUBLICAÇÃO

2005-12

RESUMO

A boron-doped diamond electrode is used for determination of Mn(II) in atmospheric particulate matter by square wave cathodic stripping voltammetry. The analytical curve was linear for Mn(II) concentrations between 5.0 and 37.5 µg L-1, with quantification limit of 3.6 µg L-1. The precision was evaluated by the relative standard deviation, with values between 5.1% and 9.3%. The electrode is free of adsorption, minimizing memory effects. Samples collected in the workplace atmosphere of a foundry had Mn(II) concentrations between 0.4 and 4 µg m-3. No significant differences were observed between the proposed method and inductively coupled plasma optical emission spectroscopy.