Comparison of USEPA 3050B and ISO 14869-1: 2001 digestion methods for sediment analysis by using FAAS and ICP-OES quantification techniques
AUTOR(ES)
Navarro, Elvira Maria Galí, Tagle, Margarita Edelia Villanueva, Marín, Maria Teresa Larrea, Alfonso, Mario Simeón Pomares
FONTE
Química Nova
DATA DE PUBLICAÇÃO
2011
RESUMO
A study of the partial USEPA 3050B and total ISO 14869-1:2001 digestion methods of sediments was performed. USEPA 3050B was recommended as the simpler method with less operational risk. However, the extraction ability of the method should be taken in account for the best environmental interpretation of the results. FAAS was used to quantify metal concentrations in sediment solutions. The alternative use of ICP-OES quantification should be conditioned by a previous detailed investigation and eventual correction of the matrix effect. For the first time, the EID method was employed for the detection and correction of the matrix effect in sediment ICP-OES analysis. Finally, some considerations were made about the level of metal contamination in the area under study.
Documentos Relacionados
- Determination of Boron in Silicon Carbide by ICP-OES and ICP-MS after Sample Preparation using Pyrohydrolysis
- Estudio preliminar sobre la determinación de elementos traza en cervezas venezolanas por ICP-OES
- Efeitos matriciais e distribuição espacial de linhas de emissão em ICP-OES com configuração axial.
- DETERMINAÇÃO SIMULTÂNEA DE METAIS E ENXOFRE POR ICP-OES EM RESÍDUOS DE DESTILAÇÃO DE PETRÓLEO
- The routine development to determination of Rare Earth elements in geological materials by ICP-OES and ICP-MS. Aplictaion in granitoid rocks from Piedade-Ibiúna (SP) and Cunhaporanga (PR) complexes