Comparison of the Beckman Auto ICS and the Syva Autolab 6000 for determination of gentamicin levels in serum.
AUTOR(ES)
Glew, R H
RESUMO
Two fully automated drug-monitoring systems, the Syva Autolab 6000 and the Beckman Auto ICS, were compared in terms of accuracy, precision, speed of operation, and cost effectiveness in the determination of gentamicin levels in serum. Within-run and between-run precision for both systems were acceptable (coefficient of variation, 2.0 to 6.9%), and patient sample comparisons resulted in an intermethod correlation coefficient of 0.96. When reference samples (prepared to contain 1.2 to 10 micrograms of gentamicin per ml) were assayed, the Syva Autolab 6000 obtained concentrations within 8% of expected values, whereas the Beckman Auto ICS reported values up to 17% higher than target values. In a time and cost comparison, reagent costs for the Beckman system were ca. 50% less than for the Syva Autolab 6000; the Syva system, however, determined patient results two to three times faster than the Beckman Auto ICS.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=271569Documentos Relacionados
- Comparison of enzyme-multiplied immunoassay technique with fluorescence polarization immunoassay for determination of gentamicin and tobramycin levels in serum.
- Rate nephelometric determination of rheumatoid factor: comparison between Kallestad QM-300 and Beckman ICS-II (RF) methods.
- Comparison of radioimmunoassay and fluorescent polarization immunoassay for quantitative determination of vancomycin concentrations in serum.
- Comparative evaluation of three methods for measuring gentamicin and tobramycin in serum.
- Evaluation of the macro-vue latex agglutination test for quantitation of gentamicin in human serum.