Comparison of BD Phoenix to Vitek 2, MicroScan MICroSTREP, and Etest for Antimicrobial Susceptibility Testing of Streptococcus pneumoniae▿

AUTOR(ES)
FONTE

American Society for Microbiology (ASM)

RESUMO

The performance of the BD Phoenix Automated Microbiology System (BD Diagnostic Systems) was compared to those of the Vitek 2 (bioMérieux), the MicroScan MICroSTREP plus (Siemens), and Etest (bioMérieux) for antibiotic susceptibility tests (AST) of 311 clinical isolates of Streptococcus pneumoniae. The overall essential agreement (EA) between each test system and the reference microdilution broth reference method for S. pneumoniae AST results was >95%. For Phoenix, the EAs of individual antimicrobial agents ranged from 90.4% (clindamycin) to 100% (vancomycin and gatifloxacin). The categorical agreements (CA) of Phoenix, Vitek 2, MicroScan, and Etest for penicillin were 95.5%, 94.2%, 98.7%, and 97.7%, respectively. The overall CA for Phoenix was 99.3% (1 very major error [VME] and 29 minor errors [mEs]), that for Vitek 2 was 98.8% (7 VMEs and 28 mEs), and those for MicroScan and Etest were 99.5% each (19 and 13 mEs, respectively). The average times to results for Phoenix, Vitek 2, and the manual methods were 12.1 h, 9.8 h, and 24 h, respectively. From these data, the Phoenix AST results demonstrated a high degree of agreement with all systems evaluated, although fewer VMEs were observed with the Phoenix than with the Vitek 2. Overall, both automated systems provided reliable AST results for the S. pneumoniae-antibiotic combinations in half the time required for the manual methods, rendering them more suitable for the demands of expedited reporting in the clinical setting.