Commissioning of the injector system of the IFUSP Microtron / "Início de operação e caracterização do sistema injetor do Mícrotron do IFUSP"
AUTOR(ES)
Tiago Fiorini da Silva
DATA DE PUBLICAÇÃO
2006
RESUMO
In this work we present the commissioning of the IFUSP Microtron injector system. We developed a new method to treat misalignments on a single lens as well as in a group of them. We installed an image acquisition system to acquire beam images from the fluorescent screen monitor. Emittance was measured and found to be (2,32 ± 0,05) pi.mm.mrad, independently of the beam energy, showing the limitation imposed by the collimator placed at the entrance of the chopper cavity.
ASSUNTO(S)
beam optics Óptica eletrônica nuclear physics física nuclear aceleração de partículas particle acceleration