Assay of netilmicin, using enzyme immunoassay for gentamicin.
AUTOR(ES)
Larson, T
RESUMO
A homogenous enzyme immunoassay for the quantitative determination of netilmicin in serum was developed. The procedure utilizes a commercially available assay for gentamicin (EMIT; Syva Co., Palo Alto, Calif.). The method was adapted to a microcentrifugal analyzer, and log-logit regression analysis was performed with a computer. The results of samples assayed by this method correlate well with microbioassay (r2 = 0.985) and radioimmunoassay (r2 = 0.986). This method is not only precise and accurate, but also very rapid and economical and compares favorably to other available methods of netilmicin assay.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=181903Documentos Relacionados
- Rapid, reproducible enzyme immunoassay for gentamicin.
- Evaluation of an enzyme immunoassay for serum gentamicin.
- Performance characteristics of bioassay, radioenzymatic assay, homogeneous enzyme immunoassay, and high-performance liquid chromatographic determination of serum gentamicin.
- In Vitro Activity of Netilmicin, Gentamicin, and Amikacin
- Homogeneous enzyme immunoassay for netilmicin.