An in-vitro morphological study of Q-switched neodymium/YAG laser trabeculotomy.
AUTOR(ES)
Venkatesh, S
RESUMO
Laser trabeculotomies produced by directing a pulsed neodymium/YAG laser beam at specimens of human anterior chamber angle obtained post mortem or after enucleation were studied by light microscopy and by scanning and transmission electron microscopy to assess the dimensions of the openings created in the trabecular meshwork, their penetrance to the canal of Schlemm, and the extent or absence of laser induced cellular damage in immediately adjacent tissue. A pulse duration of 40-50 ns at energy levels of around 30 mJ was used and the laser cavity carefully tuned to give a Gaussian spatial mode pattern. Openings in the trabecular meshwork typically of 100 microns in diameter and penetrating through to the canal of Schlemm could be regularly created with only minimal damage to adjacent tissue as judged by transmission electron microscopy. The information so gained may be useful in determining the parameters required for successful laser trabeculotomy as a treatment for primary open-angle glaucoma.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=1040926Documentos Relacionados
- Parameters for neodymium-YAG laser trabeculotomy: an in-vitro study.
- In vitro studies with a pulsed neodymium/YAG laser.
- Pulsed neodymium-YAG laser trabeculotomy: energy requirements and replicability.
- Beam obstruction hazard on Q-switched Nd-YAG ophthalmic lasers: a safety device for the Nanolas laser.
- Neodymium-YAG laser iridotomy in angle closure glaucoma: preliminary study.