A technique for anodally blocking large nerve fibres through chronically implanted electrodes.
AUTOR(ES)
Brindley, G S
RESUMO
If a spinal root of a baboon or rhesus monkey is trapped in an initially loose-fitting silicone rubber channel containing two or more platinum electrodes, electrical pulses sent through these electrodes can stimulate nerve fibres close to the cathode and block the resulting impulses close to the anode. We show (1) how anodal break excitation and excitation of fibres outside the silicone rubber channel can be avoided; (2) that an implant 26 months old behaves like a recent one; (3) that in a root containing somatic motor fibres and parasympathetic fibres, all somatic motor fibres can be blocked and most or all parasympathetic fibres excited but not blocked; (4) that provided that the electrodes pass no net direct current, prolonged stimulation with block can be harmless; (5) how block can be achieved in one direction only along a root; (6) that a peripheral nerve can be blocked by the same techniques.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=490780Documentos Relacionados
- Major venous thrombosis: a complication of transvenous pacemaker electrodes.
- Dielectrophoretic manipulation of cells with spiral electrodes.
- Bacterial and fungal killing by iontophoresis with long-lived electrodes.
- Single-Unit Stability Using Chronically Implanted Multielectrode Arrays
- Transient corneal changes associated with the use of gold foil electrodes.