A report on the interlaboratory quantitation of haemoglobin A2 and haemoglobin F
AUTOR(ES)
White, J. M.
RESUMO
An interlaboratory trial of the quantitation of Hb A2 and Hb F has been carried out on two samples of blood by 90 laboratories in the British Isles. The results have been compared with those obtained by four reference laboratories. Overall, the correlation was very poor and indicated that the methods used have to be standardized before interlaboratory values become meaningful. Moreover, in many laboratories the level of Hb A2 in a patient with proven beta thalassaemia was not reported as being increased.
ACESSO AO ARTIGO
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=477904Documentos Relacionados
- Haemoglobin A2 level in pregnancy
- Quantitative determination of haemoglobin A2 using paper electrophoresis